Ultraviolet–visible spectroscopy

physical science Maturity 11-13 Next article →

Ultraviolet–visible spectroscopy, often called UV–Vis, is a powerful method used to study how light interacts with matter.

DU640 spectrophotometer
DU640 spectrophotometer

This technique focuses on the ultraviolet and visible regions of the electromagnetic spectrum. Scientists use it to identify and quantify specific chemical compounds in many different samples. It is a very popular method because it is relatively inexpensive and easy to use. For a substance to be studied this way, it must contain a chromophore. A chromophore is a specific part of a molecule that is capable of absorbing light energy.

The mechanism of UV–Vis spectroscopy relies on the movement of electrons within a molecule. When a photon, which is a tiny particle of light, hits a chromophore, it provides energy. This energy causes an electron to jump from its current position to a higher energy molecular orbital. This process creates what scientists call an excited state. In organic molecules, there are four common types of these electronic transitions: π–π*, n–π*, σ–σ*, and n–σ*. Transition metal complexes also show interesting behavior because they have incompletely filled d orbitals. These unfilled orbitals allow them to absorb visible light, which is why many metal complexes appear colored.

Simplified UV-vis diagram
Simplified UV-vis diagram

Scientists measure several different parameters during these experiments. They might measure absorbance (A), which is the amount of light soaked up by the sample. They can also measure transmittance (%T), which is the percentage of light that passes through. Another option is reflectance (%R), which measures light that bounces off the surface. A UV–Vis spectrophotometer works by passing a beam of light through a sample. The instrument then measures how much light is absorbed at every specific wavelength. This data is often presented as a spectrum, showing peaks where absorption is strongest.

A fundamental principle in this field is the Beer–Lambert law. This law provides a mathematical way to link light absorption to the properties of a solution. It states that absorbance is directly proportional to the concentration of the absorbing species. It also depends on the path length, which is the distance the light travels through the sample. The formula is written as A = εLc. Here, ε represents the molar absorptivity, or extinction coefficient. This coefficient is a constant that describes how strongly a substance absorbs light at a specific wavelength. It is a fundamental property that depends on the solvent, temperature, and pressure.

Bis(triphenylphosphine) nickel (II) chloride UV-vis
Bis(triphenylphosphine) nickel (II) chloride UV-vis

Because of this law, UV–Vis is widely used for quantitative analysis. This means scientists use it to find out exactly how much of a substance is present. For example, they can determine the concentration of transition metal ions or biological macromolecules. They can even monitor structural changes in DNA. In the semiconductor industry, the technology is used to measure the thickness of thin films on a wafer. To ensure accuracy, scientists often use a calibration curve. This involves comparing the response of an unknown sample to a known standard.

However, many variables can influence the results of a measurement. The type of solvent used can change the absorption spectrum significantly. For instance, water is often used for water-soluble compounds, while ethanol is used for organic-soluble ones. The pH level and the temperature of the solution also matter. Even the spectral bandwidth of the instrument can change the outcome. Bandwidth is the range of wavelengths the instrument transmits at one time. A narrow bandwidth provides higher resolution and better accuracy. A wider bandwidth allows for faster scanning but may result in lower resolution.

Bis(triphenylphosphine) nickel (II) chloride UV-vis
Bis(triphenylphosphine) nickel (II) chloride UV-vis

Scientists must also watch out for a problem called stray light. Stray light is any light that reaches the detector that was not the specific wavelength selected. This light can be caused by reflections or scattering inside the instrument. Stray light can cause significant errors because it makes the reported absorbance appear lower than it actually is. To prevent this, some scientists use a double monochromator. A single monochromator might only allow measurements up to about 2 Absorbance Units (AU) before stray light becomes a problem. A double monochromator can handle much higher levels, up to about 6 AU.

UV–Vis spectroscopy connects many different scientific disciplines. It is a routine tool in analytical chemistry for identifying functional groups in molecules. Scientists can use the Woodward–Fieser rules to predict the wavelength of maximum absorption, known as λmax, for certain organic compounds. It is also used in biochemistry to study large molecules. By understanding how light and electrons interact, researchers can gain deep insights into the chemical makeup of our world.

752 words

Where will your curiosity go next?

Keep exploring at your chosen reading level.

🖼️ Images & Media (3)
File:DU640 spectrophotometer.jpg
DU640 spectrophotometer.jpg
File:Bis(triphenylphosphine) nickel (II) chloride UV-vis.JPG
Bis(triphenylphosphine) nickel (II)...